StatusCommandGroup
Bases: ModuleCommandGroupBase
A command group for getting the status of the probe station and controlling the dashboard module.
get_access_level()
Retrieves the access level of operation.
Returns:
| Type | Description |
|---|---|
AccessLevel
|
A string representing the access level. Possible values are: |
AccessLevel
|
'Operator', 'Admin', 'Service', 'Engineer', 'Debug'. |
get_chuck_temp()
Get current chuck temperature.
Returns:
| Type | Description |
|---|---|
float
|
The chuck temperature in degrees Celsius. |
get_chuck_temp_setpoint()
Get current chuck temperature setpoint.
Returns:
| Type | Description |
|---|---|
float
|
The chuck temperature setpoint in degrees Celsius. |
get_chuck_thermo_energy_mode()
Get the current chuck thermo energy mode.
Returns:
| Type | Description |
|---|---|
str
|
The current energy mode as a string. Possible values: Fast, Optimal, HighPower, Customized. |
get_chuck_thermo_hold_mode()
Get thermo chuck hold mode.
Returns:
| Type | Description |
|---|---|
str
|
The current hold mode. Possible values: Active, Nonactive. |
get_chuck_thermo_state()
Return thermo chuck state.
Returns:
| Type | Description |
|---|---|
ThermoChuckState
|
A tuple with the current state of the thermo chuck. Contains six boolean variables: |
ThermoChuckState
|
isCooling, isHeating, isControlling, isStandby, isError, isUncontrolled. |
get_high_purge_state()
Get thermo chuck high purge state.
Returns:
| Type | Description |
|---|---|
str
|
The current high purge state. Possible values: ON, OFF. |
get_machine_id()
Retrieves the machine ID.
Returns:
| Type | Description |
|---|---|
str
|
A string containing the machine ID. |
get_machine_status()
Get machine status.
Returns:
| Type | Description |
|---|---|
Tuple[bool, bool, bool]
|
A tuple with the current status of the machine. Contains three boolean variables: isInitialized, isMeasuring, LoaderBusy. |
get_soaking_time(temperature)
Get the thermochuck soaking time in seconds that is set up for a certain temperature in the dashboard.
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
temperature |
float
|
The temperature. This temperature value must be one of the predefined temperature values set up in the dashboard. |
required |
Raises:
| Type | Description |
|---|---|
ProberException
|
With error code 200 is thrown when the temperature is not a predefined value set up in the dashboard. |
Returns:
| Type | Description |
|---|---|
|
The soaking time in seconds. |
get_version()
Retrieves the system version.
Returns:
| Type | Description |
|---|---|
str
|
A string containing version information. |
set_chuck_temp(temp, lift_chuck=False)
Set chuck temperature setpoint.
This function wraps SENTIO's status:set_chuck_temp remote command.
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
temp |
float
|
The chuck temperature setpoint in degrees Celsius. |
required |
lift_chuck |
bool
|
If True, the chuck will moved to the lift position if required. If this value is false and a move to lift is required an error occurs. |
False
|
Raises:
| Type | Description |
|---|---|
ProberException
|
If an error occurred. |
set_chuck_thermo_energy_mode(mode)
Set chuck thermo energy mode.
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
mode |
str
|
The desired energy mode. Possible values: Fast, Optimal, HighPower, Customized. |
required |
Returns:
| Type | Description |
|---|---|
None
|
None |
Raises:
| Type | Description |
|---|---|
ValueError
|
If the provided mode is not valid. |
set_chuck_thermo_hold_mode(mode)
Set thermo chuck hold mode.
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
mode |
bool
|
A boolean indicating whether to enable (True) or disable (False) hold mode. |
required |
Returns:
| Type | Description |
|---|---|
None
|
None |
set_chuck_thermo_mode(mode)
Set chuck thermo operation mode.
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
mode |
str
|
The operation mode to set. Possible values: Normal, Standby, Defrost, Purge, Turbo, Eco. |
required |
Returns:
| Type | Description |
|---|---|
None
|
None |
Raises:
| Type | Description |
|---|---|
ValueError
|
If the provided mode is not valid. |
set_high_purge(enable)
Set thermo chuck high purge state.
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
enable |
bool
|
A boolean indicating whether to enable (True) or disable (False) high purge |
required |
Returns:
| Type | Description |
|---|---|
None
|
None |
show_message(message, button=DialogButtons.Ok, caption='None', level='Hint')
Show a message dialog for user interaction.
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
message |
str
|
The message to show. |
required |
button |
DialogButtons
|
The button configuration (e.g., 'Ok', 'OkCancel', 'YesNo'). |
Ok
|
caption |
str
|
Caption for the dialog. |
'None'
|
level |
str
|
Level of the message (e.g., 'Hint', 'Warning', 'Error'). |
'Hint'
|
Returns:
| Type | Description |
|---|---|
str
|
The button that was pressed. |
start_show_message(message, button=DialogButtons.Ok, caption='None', level='Hint')
Start an asynchronous message dialog.
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
message |
str
|
The message to show. |
required |
button |
DialogButtons
|
The button configuration. |
Ok
|
caption |
str
|
The dialog caption. |
'None'
|
level |
str
|
The level of message importance. |
'Hint'
|
Returns:
| Type | Description |
|---|---|
Response
|
A string containing Command ID and button pressed info. |
handler: python options: members: StatusCommandGroup