Skip to content

AuxCommandGroup

Bases: ModuleCommandGroupBase

This command group contains functions for working with auxiliary sites of the chuck. You are not meant to create instances of this class on your own. Instead, use the aux property of the SentioProber class.

Attributes:

Name Type Description
cleaning AuxCleaningGroup

A subgroup to provide logic for probe cleaning.

get_element_info(element_standard_id, site=None)

Retrieves information of a calibration element on a calibration substrate placed on the chuck.

Wraps SENTIO's "aux:get_element_info" remote command.

According to the SENTIO specification: - If a site is given, the command format is: aux:get_element_info , - If no site is given, the command format is: aux:get_element_info

Parameters:

Name Type Description Default
element_standard_id str

The standard ID of the element (e.g. "0102").

required
site Optional[ChuckSite]

(Optional) The auxiliary site (e.g. ChuckSite.AuxRight). If omitted, the currently active site is used.

None

Returns:

Type Description
ElementInfo

An ElementInfo object containing: - element_type (ElementType) - element_subtype (str) - x_position (float) - y_position (float) - spacing (float) - touch_count (int) - life_time (float)

get_element_life_time(element_standard_id, site=None)

Retrieves the life time (in %) of an element on a calibration substrate placed on the chuck.

Wraps SENTIO's "aux:get_element_life_time" remote command.

Parameters:

Name Type Description Default
element_standard_id str

The standard ID of the element (e.g. "0102").

required
site Optional[ChuckSite]

(Optional) The auxiliary site (e.g. ChuckSite.AuxRight). If omitted, the currently active site is used.

None

Returns:

Type Description
float

The life time percentage (float).

get_element_pos(element_standard_id, site=None)

Retrieves the (X, Y) position of an element on a calibration substrate placed on the chuck.

Wraps SENTIO's "aux:get_element_pos" remote command.

Parameters:

Name Type Description Default
element_standard_id str

The standard ID of the element (e.g. "0102").

required
site Optional[ChuckSite]

(Optional) The auxiliary site (e.g. ChuckSite.AuxRight). If omitted, the currently active site is used.

None

Returns:

Type Description
Tuple[float, float]

A tuple (x_position, y_position) in micrometers (floats).

get_element_spacing(element_standard_id, site=None)

Retrieves the spacing value of an element on a calibration substrate placed on the chuck.

Wraps SENTIO's "aux:get_element_spacing" remote command.

Parameters:

Name Type Description Default
element_standard_id str

The standard ID of the element (e.g. "0102").

required
site Optional[ChuckSite]

(Optional) The auxiliary site (e.g. ChuckSite.AuxRight). If omitted, the currently active site is used.

None

Returns:

Type Description
float

The spacing in micrometers (float).

get_element_touch_count(element_standard_id, site=None)

Retrieves the touch count of an element on a calibration substrate placed on the chuck.

Wraps SENTIO's "aux:get_element_touch_count" remote command.

Parameters:

Name Type Description Default
element_standard_id str

The standard ID of the element (e.g. "0102").

required
site Optional[ChuckSite]

(Optional) The auxiliary site (e.g. ChuckSite.AuxRight). If omitted, the currently active site is used.

None

Returns:

Type Description
int

The touch count (int) of the element.

get_element_type(element_standard_id, site=None)

Retrieves the type of an element on a calibration substrate placed on the chuck.

Wraps SENTIO's "aux:get_element_type" remote command.

Parameters:

Name Type Description Default
element_standard_id str

The standard ID of the element (e.g. "0102").

required
site Optional[ChuckSite]

(Optional) The auxiliary site (e.g. ChuckSite.AuxRight). If omitted, the currently active site is used.

None

Returns:

Type Description
ElementType

An ElementType enumerator (e.g. ElementType.Short, ElementType.Thru, ElementType.Align, etc).

get_substrate_info(site)

Retrieves information of a calibration substrate or cleaning pad placed on a chuck site.

Wraps SENTIO's "aux:get_substrate_info" remote command.

Parameters:

Name Type Description Default
site ChuckSite

The auxiliary site (e.g. ChuckSite.AuxRight).

required

Returns:

Type Description
Tuple[str, str, float]

A tuple containing: - substrate_type (str) - substrate_id (str) [if the id is "OK", an empty string is returned] - life_time (float) in %

get_substrate_type(site=None)

Retrieves the type of a calibration substrate placed on the chuck.

Wraps SENTIO's "aux:get_substrate_type" remote command.

Parameters:

Name Type Description Default
site Optional[ChuckSite]

(Optional) The auxiliary site (e.g. ChuckSite.AuxRight). If omitted, the currently active site is used.

None

Returns:

Type Description
str

A string describing the substrate type.

str

If the remote command returns "OK", an empty string is returned.

retrieve_substrate_data(site=None)

Retrieves contact and home information from the configuration file and assigns it to the corresponding auxiliary site. Must be used after system start, since data are not retrieved automatically for safety reasons.

Wraps SENTIO's "aux:retrieve_substrate_data" remote command.

Parameters:

Name Type Description Default
site Optional[ChuckSite]

(Optional) The auxiliary site (e.g. ChuckSite.AuxRight). If omitted, data for all auxiliary sites is retrieved.

None

Returns:

Type Description
List[ChuckSite]

A list of ChuckSite for which data was retrieved.

step_to_dut_element(dut_name, move_z=True, x=None, y=None)

Steps to a DUT element on the currently active chuck site.

Wraps SENTIO's "aux:step_to_dut_element" remote command.

Parameters:

Name Type Description Default
dut_name str

The name of the DUT element (e.g. "RefDUT").

required
move_z bool

If True, move Z automatically. If False, skip Z movement.

True
x Optional[float]

(Optional) X position [um].

None
y Optional[float]

(Optional) Y position [um].

None

step_to_element(element_standard_id, offset_x=0.0, offset_y=0.0, motorized_positioner_move=True)

Steps to a calibration element on the currently active chuck site.

Wraps SENTIO's "aux:step_to_element" remote command.

Parameters:

Name Type Description Default
element_standard_id str

The standard ID of the element (e.g. "0102").

required
offset_x float

An X offset (in micrometers).

0.0
offset_y float

A Y offset (in micrometers).

0.0
motorized_positioner_move bool

If True, move with motorized positioner. If False, only chuck move.

True

AuxCleaningGroup

Bases: CommandGroupBase

This command group provides functions for for cleaning the probe. You are not meant to create instances of this class on your own. Instead use the cleaning property of the AuxCommandGroup.

Example:

    from sentio_prober_control.Sentio.ProberSentio import SentioProber

    prober = SentioProber.create_prober("tcpip", "127.0.0.1:35555")
    prober.aux.cleaning.enable_auto(True)

enable_auto(stat)

Enable automatic probe cleaning.

Parameters:

Name Type Description Default
stat bool

A flag indicating whether to enable or disable cleaning.

required

reset_touch_count(chuckSite)

Resets the touchdown counter of a cleaning pad.

Parameters:

Name Type Description Default
chuckSite enum ChuckSite

The chuck site the cleaning pad is placed on

required

start(touchdowns=None)

Start the cleaning procedure.

Parameters:

Name Type Description Default
touchdowns int

The number of touchdowns to perform.

None

handler: python options: members: AuxCleaningGroup